Health physics : Wide area reference sources

 

Tritium sources

These wide area tritium sources are designed for the calibration and routine checking of tritium surface contamination monitoring equipment.
Due to the extremely short range of Tritium (H-3) beta particles in air (approximately 4mm), the direct measurement of H-3 must be undertaken with great care.
Therefore it is important to check the correct function of the detector using a suitable H-3 source immediately prior to measurement.

These tritium sources are electrically conducting and do not suffer from the build up of electrostatic charge that can cause measurement problems. The thickness of the activated layer is approximately 3 microns. Tritium wide area reference sources are calibrated for β -emission only, not for activity.

Nuclide

 Nominal activity 

 Drawing 

 Active 
 dimensions 
(mm)

 Overall 
 dimensions 
(mm)

 Product Code 

Remark

H-3

10 kBq

VZ-1610

10 x 75

134 x 210

TRRB1951

for Eberline tritium contamination monitor FHT111M
ß-emission rate 100 - 300 1/s

H-3

10 kBq

VZ-1516

15 x 152

86 x 226

TRR16161

for Berthold tritium contamination monitor
ß-emission rate 100 - 300 1/s


If your required product is not shown, or you cannot find the details you are searching for, Entsys Technologies will try to assist you. Please email us at entsys@anet.net.th.