DOSE RATE METER 6150AD AND ACCESSORIES |
TECHNICAL INFORMATION CONCERNING
THE DOSE RATE METER 6150AD
The instrument has a built-in GM
counter and additionally provides a probe socket to
connect external probes. If a probe is connected, the
internal counting tube is automatically turned off; the
detector of the probe is now measuring. The type of the
probe is automatically recognized (e.g. display "ext 17"
when a 6150AD-17 probe is connected). Display range and
units are chosen automatically according to the type of
the probe. Also, there is no fixed assignment between
probe and instrument. That means you may connect any
probe with any 6150AD, where probe indication will not
depend on the 6150AD in use. One pin of the probe socket
serves as serial interface and can be used to transmit
and evaluate the display (dose rate) of the instrument
on a PC. Cables to connect external probes may be up to
100 m long (depending on probe type). Standard lengths
available from stock are 1.25 m, 3 m, 10 m, 20 m, 75 m,
and 100 m.
|
| Model
summary
The basic 6150AD models
are 6150AD1 to 6150AD6, henceforth generally designated
as "6150ADn" (where n stands for 1 to 6).
Models 6150AD1/3/5 have
the same range of up to 1 Sv/h. They only differ in
number and extent of functions, that is in software
(EPROM). The 6150AD3 was the more intelligent successor
of the 6150AD1. The 6150AD5 is again the successor of
the 6150AD3 with additional support for the scintillator
probe 6150AD-b. Likewise, models 6150AD2/4/6 have a
range of up to 10 mSv/h, where the 6150AD4 was the more
intelligent successor of the 6150AD2, and the 6150AD6
again corresponds to a 6150AD4 with additional support
for the scintillator probe. In that sense models 6150AD1
to 6150AD4 are outdated (but are still available,
particularly the models 6150AD1 and 6150AD2 for those
users who prefer easier operation of a simple model).
The following Table 1 summarizes all models 6150AD1 to
6150AD6 with their essential features.
Table
1
6150AD
model summary
|
range up
to
|
additional
functions
|
supports
probe
6150AD-b
|
| 6150AD1 |
1 Sv/h
|
no
|
no
|
| 6150AD3 |
yes
|
| 6150AD5 |
yes
|
| 6150AD2 |
10
mSv/h
|
no
|
no
|
| 6150AD4 |
yes
|
| 6150AD6 |
yes
|
The additional functions comprise:
protection against unintentionally switching off,
non-volatile storage of the dose measured with the
internal tube, dose alarm, programmable alarm thresholds
for dose and dose rate, better resolution (more places
after the decimal point) when indicating dose, alarm
threshold, and so on.
|
| Radiation
quantity
Unless stated otherwise,
6150AD and probes are designed for photon dose
equivalent (rate) Hx. Models designed for ambient dose
equivalent (rate) H*(10) get the suffix "/H" or "/E" to
their type name. The only difference between "/H" and
"/E" models is their compatibility with probes (seeTable
2 below). "/E" models of the probes and "/E" models of
the 6150AD only work among each other. "/H" models work
among each other and with Hx models. For example, a
6150AD5 works with a probe 6150AD-18/H. Using that
6150AD5 without a probe yields Hx, whereas connecting
the 6150AD-18/H probe yields H*(10). The other way
round, a 6150AD5/H works with a probe 6150AD-18. In that
case, the 6150AD/H yields H*(10), whereas connecting the
6150AD-18 probe yields Hx. "/H" models of probes and
meter offer the advantage that they remain compatible
with older Hx meters and probes. "/E" models are
recommended in case you want to force the 6150AD's
indication to H*(10) even when working with external
probes. However, this will require your possibly
existing Hx equipment to be upgraded to "/E" models if
it shall be used together with the new "/E" equipment.
Note that the difference between Hx and H*(10) is not
very great except at low photon
energies.
Table
2
Compatibility of 6150AD meter and
probes
| Probe |
Meter
|
6150ADn,
6150ADn/H
|
6150ADn/E
|
Sievert
probes
neither /H nor /E models
(i.e., for Hx) |
yes
|
no
|
Sievert
probes
/H
models (i.e., for H*(10)) |
yes
|
no
|
Sievert
probes
/E
models (i.e., for H*(10)) |
no
|
yes
|
| Pulse rate probes
(s-1) |
yes
|
yes
|
|
DOSE RATE METER 6150AD5
For photon radiation (gamma- and
X-radiation), with analog and digital display, automatic
range switching, non-volatile dose memory, individually
programmable dose and dose rate alarm thresholds,
indication of relative standard deviation, including
9V-battery (IEC 6LR61).
Analog range: 1 µSv/h - 1000 mSv/h
Digital range: 0.1 µSv/h - 999
mSv/h
Energy range: 45 keV -
3 MeV
- Hx model (compatible
with all probes except /E models): 6150AD5
- H*(10) model (compatible with all
probes except /E models): 6150AD5/H
- H*(10) model (compatible with
pulse rate probes and /E probes): 6150AD5/E
|
| DOSE RATE METER
6150AD6
for photon radiation (gamma- and
X-radiation), with analog and digital display, automatic
range switching, non-volatile dose memory, individually
programmable dose and dose rate alarm thresholds,
indication of relative standard deviation, including
9V-battery (IEC 6LR61).
Analog range: 0.1 µSv/h - 10 mSv/h
Digital range: 0.01 µSv/h - 9.99
mSv/h
Energy range: 60 keV -
1.3 MeV
- Hx model
(compatible with all probes except /E models): 6150AD6
- H*(10)
model (compatible with all probes except /E models): 6150AD6/H
- H*(10)
model (compatible with pulse rate probes and /E probes): 6150AD6/E
|
| GAMMA PROBE 6150AD-15
High dose rate probe for photon
radiation (gamma- and X-radiation), power supply from
6150AD.
Analog range: 0.1
mSv/h - 10 Sv/h
Digital
range: 0.01 mSv/h - 9.99 Sv/h
Energy range: 65 keV - 3 MeV
- Hx model (compatible with all meters
6150ADn and 6150ADn/H): 6150AD-15
- H*(10) model (compatible with all
meters 6150ADn and 6150ADn/H): 6150AD-15/H
- H*(10) model (compatible with
6150ADn/E meters only): 6150AD-15/E |
| GAMMA PROBE
6150AD-18
Low dose rate probe for photon
radiation (gamma- and X-radiation), power supply from
6150AD.
Analog range: 0.1
µSv/h - 10 mSv/h
Digital
range: 0.01 µSv/h - 9.99 mSv/h
Energy range: 65 keV - 1.3 MeV
- Hx model (compatible with all
meters 6150ADn and 6150ADn/H): 6150AD-18
- H*(10) model (compatible with all
meters 6150ADn and 6150ADn/H): 6150AD-18/H
- H*(10) model (compatible with
6150ADn/E meters only): 6150AD-18/E
|
SCINTILLATOR PROBE 6150AD-b
For photon
radiation (gamma- and X-radiation), plastic scintillator
3" x 3", power supply from 6150AD, including helical
probe cable.
Analog range:
10 nSv/h - 100 µSv/h
Digital
range: 1 nSv/h - 99.99 µSv/h
Energy range 6150AD-b: 23 keV - 7 MeV
Energy range 6150AD-b/H, /E: 20 keV
- 7 MeV
- Hx model
(compatible with 6150AD5(/H) and 6150AD6(/H) meters): 6150AD-b
- H*(10)
model (compatible with 6150AD5(/H) and 6150AD6(/H)
meters): 6150AD-b/H
-
H*(10) model (compatible with 6150AD5/E and 6150AD6/E
meters only): 6150AD-b/E
|
| ALPHA-BETA-GAMMA
PROBE 6150AD-17
Wipe test probe, compatible with any
Dose Rate Meter 6150AD, power supply from 6150AD.
Analog range: 0.1 s-1 -
10k s-1
Digital
range: 0.01 s-1 - 9.99k s-1
|
| BETA-GAMMA LIQUID-SAMPLE PROBE
6150AD-19
for liquids, compatible with any
Dose Rate Meter 6150AD, power supply from 6150AD.
Analog range: 0.1 s-1 -
10k s-1
Digital
range: 0.01 s-1 - 9.99k s-1
|
CONTAMINATION PROBE 6150AD-k
large area counting tube for alpha-,
beta- and gamma-radiation, compatible with any Dose Rate
Meter 6150AD, power supply from 6150AD. Including fixed
probe cable.
Analog range:
0.1 s-1 - approx. 40k s-1
Digital range: 0.01 s-1 - approx. 40k s-1
|
TELETECTOR PROBE 6150AD-t
for beta radiation and photon
radiation (gamma- and X-radiation). Two counting tubes,
auto-ranging between low dose rate and high dose rate
counting tube. Telescope extendible to a length of
approx. 4 m. Power supply from 6150AD. Including fixed
probe cable.
Analog range:
0.1 µSv/h - 10 Sv/h
Digital
range: 0.01 µSv/h - 9.99 Sv/h
Energy ranges:
low dose rate counting tube: 65 keV - 1.3
MeV
high dose rate counting
tube: 65 keV - 3 MeV
- Hx
model (compatible with all meters 6150ADn and
6150ADn/H): 6150AD-t
- H*(10) model (compatible with all meters
6150ADn and 6150ADn/H): 6150AD-t/H
- H*(10) model (compatible with
6150ADn/E meters only): 6150AD-t/E
|
GAMMA ALARM
STATION 859.1
for storage of one Dose Rate Meter
6150AD, with particularly well noticeable audible and
optical alarm when exceeding the adjustable dose rate
alarm threshold. Dose rate measurements by 6150AD or
probe connected to 6150AD (6150AD not included).
Suitable for wall mounting, including lead accumulator,
power supply 87-250VAC / 50 Hz.
|
PROBE MULTIPLEXER 861.1
for storage of one Dose Rate Meter
6150AD with possibility to connect up to 10 probes of
the 6150AD-family (6150AD not included).
Selection of probe channel at Multiplexer or via
connected PC. Recording and graphic representation of
the channels on PC (standard software included).
Suitable for wall mounting, power supply 87-250VAC / 50
Hz.
|
|
|

 |
 |
 |